Why Two CO₂ Monitors Disagree, and How to Check Yours

You set a new CO₂ monitor next to the one you already own, and an hour later one shows 800 ppm while the other shows 900. Which one is right? Quite often both are, within what their makers promise, and the gap says more about how the sensors work than about the air in your room.

You can check a monitor at home with two simple tests: a side-by-side comparison and a few minutes in outdoor air. Together they tell you whether a disagreement matters for the one decision a CO₂ monitor is really good at, which is when to let fresh air in.

Two CO₂ monitors on an indoor shelf showing 800 and 900 beside an open window, and a third monitor on a garden table outside showing 420.
Two monitors on the same shelf rarely show the same number; a few minutes in outdoor air shows whether each one is close to the fresh-air baseline.

Key takeaways

  • Two monitors can both be right and still differ. Typical accuracy specs of ±(30 to 50 ppm + 2.5 to 5 % of the reading) allow gaps of 100 ppm or more at ordinary indoor levels.
  • Only NDIR sensors measure CO₂ itself. An “eCO₂” value is estimated from other gases, so it can’t be checked against fresh air.
  • Self-calibration needs fresh air. Automatic baseline correction assumes the sensor regularly sees outdoor air near 400 ppm; a monitor in a room that’s never aired can slowly drift low.
  • Trends guide ventilation, not the last digit. Both monitors rise when the door closes and fall when a window opens, and that’s the signal you act on.

Two kinds of sensor can show the same “ppm”

Most monitors that measure CO₂ itself use a nondispersive infrared sensor, NDIR for short. A tiny light source shines through a small chamber onto a detector tuned to a wavelength CO₂ absorbs; the more CO₂, the less light arrives. Some sensors, such as Sensirion’s SCD4x family, pick up the absorbed light as a tiny pressure pulse, a photoacoustic version of the same idea (SCD4x datasheet). Because NDIR responds to something CO₂ physically does, an NDIR reading can be compared directly with outdoor air, and it’s the type the UK workplace regulator recommends for checking ventilation (Health and Safety Executive).

Many cheaper “air quality” gadgets show eCO₂ instead. It comes from a metal-oxide sensor, a small heated surface whose electrical signal changes when gases such as hydrogen or alcohols touch it (MOX for short), and software converts that signal into a “CO₂ equivalent” (SGP30 datasheet). The idea is that people give off small amounts of such gases along with CO₂, so in an occupied room the estimate often moves in the right direction; a cleaning spray or frying onions can push it up just as well, with no change in ventilation. If one of your monitors shows eCO₂, the gap isn’t a calibration question; the two devices measure different things.

What the accuracy line on a datasheet allows

NDIR accuracy is written as a fixed part plus a percentage of the reading. Sensirion specifies its basic SCD40 to ±(50 ppm + 5 % of reading) between 400 and 2,000 ppm, and the SCD41 to ±(50 ppm + 2.5 %) up to 1,000 ppm (SCD4x datasheet). Senseair’s S8 module is specified to ±(40 ppm + 3 %) between 400 and 2,500 ppm (S8 LP specification).

At 1,000 ppm that means up to 100 ppm off for the SCD40, 75 ppm for the SCD41 and 70 ppm for the S8. Two healthy SCD40 sensors could show 900 and 1,100 ppm in the same air and both meet their spec. Senseair’s figure also applies only after three self-calibration periods of continuous operation.

One sensor agrees with itself far better than with its neighbor; Sensirion gives a typical repeatability of ±10 ppm. That’s why the trend from one monitor is more trustworthy than its absolute number. Sensors also drift slowly, and in one long-term comparison low-cost NDIR sensors left uncorrected for a year read up to about 24 ppm low against a reference instrument (Cai and colleagues, 2024).

Self-calibration assumes your monitor regularly sees fresh air

To cancel that drift, most NDIR monitors run automatic baseline correction, or ABC (Sensirion calls it ASC). The software tracks the lowest reading over a set period and assumes it was fresh outdoor air at 400 ppm. For the S8 the period is eight days, with each correction limited to about 30 to 50 ppm (S8 LP specification). The SCD4x expects 400 ppm air for more than three minutes at least once per week of operation, corrects first after 44 hours and then every 156 hours (SCD4x datasheet).

That works well where windows open every day. It works poorly in a room that never gets close to outdoor levels, such as a bedroom kept shut all winter or a room that’s almost never empty. If the lowest reading of the week is 600 ppm, the algorithm decides that was 400 ppm and pulls the readings down, step by step. Two monitors with a different fresh-air history can drift apart while both work as designed.

The 400 ppm reference also sits a little below today’s air. The global monthly mean was 424.68 ppm in August 2026 (NOAA Global Monitoring Laboratory), and city air can rise well above it, in places approaching 600 ppm for a while (ASHRAE position document).

Altitude, warm-up and the first weeks of a new monitor

An NDIR sensor effectively counts CO₂ molecules in its chamber. Higher up, the air is thinner, so an uncorrected sensor reads low. Senseair puts the effect at 1.6 % of the reading per kilopascal (10 hPa) away from normal pressure; where local pressure sits 50 hPa below the standard 1,013 hPa, an uncorrected S8 showing 1,000 ppm would be about 80 ppm low. The SCD4x accepts an altitude or a live pressure value and assumes sea level by default. If your monitor’s app asks for your altitude, enter it. Weather apps usually show pressure adjusted to sea level, which hides this effect.

Electrical warm-up is quick, under 60 seconds for the S8. Calibration warm-up isn’t: a new SCD4x makes its first self-correction after 44 hours, and the S8 accuracy figure applies after three eight-day periods. Both makers note that shipping and rough handling can shift readings, so a brand-new monitor that disagrees with an older one in its first days isn’t a surprise.

Where the monitor sits changes the number

Exhaled breath carries far more CO₂ than room air, so a monitor on the bedside table or beside your laptop partly measures you. Place it at head height, more than 50 cm from people, and away from windows, doors and air vents (Health and Safety Executive). By a tilted window it reads the incoming draft, and near a door it catches hallway air. Keep it out of direct sun and off radiators too. Before blaming calibration, put both monitors in the same spot.

A side-by-side test and an outdoor fresh-air check

The side-by-side test separates placement from calibration and takes an evening:

  • Set both monitors next to each other at the same height, more than 50 cm from anyone and away from windows and vents.
  • Let them run together for a few hours, or a night in a closed bedroom, to cover a range of levels.
  • Note both readings every 15 minutes with the time and anything that happened, like a window opened or cooking started (the measurement record guide has a simple format).
  • Read the pattern. A steady gap points to calibration; a gap that grows with the level points to a scaling or altitude setting; one jumping during cooking while the other stays calm points to eCO₂.

The outdoor check shows which monitor sits closer to the true baseline. On a mild, dry day, put the monitor in shade away from roads, chimneys and running engines, then step back so your breath doesn’t reach it. Give it 10 to 15 minutes; the sensors cover about two-thirds of a sudden change within a minute (60 seconds for the SCD4x, 40 seconds or less for the S8), but the housing takes longer to settle.

Fresh air is around 425 ppm today. With a tolerance of roughly ±55 to 70 ppm at that level, a healthy NDIR monitor in clean outdoor air usually lands between about 360 and 500 ppm. Beside a busy road, 520 ppm isn’t a failure; near 300 or 700 ppm on a calm day in a park is worth a closer look.

Many monitors offer a manual “outdoor” or 400 ppm calibration. Use it only outdoors under these steady conditions, since Sensirion asks for at least three minutes in air of constant CO₂ and Senseair’s background calibration assumes 400 ppm fresh air. Pressed indoors, the same button teaches the sensor a wrong baseline.

What a 100 ppm gap means for airing a room

For ventilation decisions, almost nothing. In a closed, occupied room CO₂ climbs by hundreds of ppm over an evening or a night (see what changes overnight CO₂ in a closed bedroom), and both monitors follow that climb. What you act on is the shape of the curve: rising when the door shuts, falling when a window opens. Indoor CO₂ is at best an indicator of how much outdoor air reaches each person (ASHRAE), and a 100 ppm offset only moves the moment you’d open the window a little earlier or later.

A gap deserves attention when it keeps growing over weeks, when one monitor fails the outdoor check by a wide margin, when indoor readings sit well below 400 ppm, or when one monitor barely drops with a window wide open. Otherwise, keep the monitor that passed the outdoor check in one good spot and use it as described in CO₂ as a ventilation compass.

FAQ

Which of my two monitors should I trust?

Trust the NDIR monitor that comes closest to outdoor air in the fresh-air check. If both pass, pick one and stay with it; one device compared with itself over time beats two compared with each other.

Why does my monitor show less than 400 ppm indoors?

An occupied room should sit at or above outdoor levels, so a reading well below 400 ppm usually means the baseline has slipped, often through self-calibration in a room that never saw fresh air or a manual calibration done indoors. An outdoor check and an outdoor recalibration usually sort it out.

How often should I repeat the outdoor check?

When a monitor is new, after moving to a different altitude, after shipping or a fall, and every few months while windows stay mostly closed. Note each result in your measurement record to see whether the monitor stays stable.